Unpacking, Installation And Commissioning; Place Of Installation, Place Of Use; Unpacking; Installation - KERN 572-30 Mode D'emploi

Masquer les pouces Voir aussi pour 572-30:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6

Unpacking, installation and commissioning

6.1 Place of installation, place of use

The balance is constructed in such a way that reliable weighing results can be achieved
under normal application conditions.
By selecting the correct location for your balance, you will be able to work quickly and pre-
cisely.
Therefore please observe the following at the place of installation:
Place the balance on a firm, level surface;
-
Avoid extreme heat as well as temperature fluctuation caused by installing next to a
-
radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
-
Avoid jarring during weighing;
-
Protect the balance against high humidity, vapours and dust;
-
Do not expose the device to extreme dampness for longer periods of time. Inadmis-
-
sible bedewing (condensation of air moisture on the device) can occur if a cold de-
vice is taken into a significantly warmer environment. In this case, please acclima-
tise the device for approx. 2 hours at room temperature after it has been discon-
nected from the mains.
Avoid static charging of the material to be weighed, weighing container and wind-
-
shield.
Major display deviations (incorrect weighing results) are possible if electromagnetic fields
occur as well as due to static charging and instable power supply. It is then necessary to
change the location.

6.2 Unpacking

Carefully remove the balance from its packaging, remove the plastic wrapping and position
the balance in its intended working location.

6.2.1 Installation

Install the balance in such a fashion that the weighing plate is absolutely horizontally.

6.3 Mains supply

Electric power supply is by means of the external mains supply circuit. The printed voltage
level must comply with the local voltage.
Only use original KERN mains supply circuits. The use of other makes is subject to ap-
proval by Kern.