Verification - KERN ABT Serie Mode D'emploi

Masquer les pouces Voir aussi pour ABT Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9 Verification

General introduction:
According to EU directive 90/384/EEC balances must be verified if they are used as
follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purposes.
d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
Verification instructions
An EU type approval exists for balances described in their technical data as verifya-
ble. If a balance is used where obligation to verify exists as described above, it must
verified and re-verified in regular intervals.
Re-verification of a balance is carried out according to the respective national regula-
tions. The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
After verification the balance is sealed at the indicated positions.
Verification of the balance is invalid without the "seal".
Position of the "official seals":
Balances with obligation to verify must be taken out of operation if:
The weighing result of the balance is outside the error limit. Therefore, in
-
regular intervals load balance with known test weight (ca. 1/3 of the max.
load) and compare with displayed value.
The reverification deadline has been exceeded.
-
30
Seals
ABT-NM-BA-e-1622

Publicité

Table des Matières
loading

Table des Matières