13 Description of individual functions
13.1 Stability filter
Generally, data processing for greater stability slows the response and processing
for higher response reduces stability. Balances of the series ABT are designed to
ensure both qualities, that is, fast reaction time and high stability.
Most measurements can be carried out using default settings, that is, standard
mode. In specific environmental conditions and for specific measuring purposes anti-
convection, high stability or apportioning mode is available for you to use. The cur-
rently set mode is shown on the setting control display (see ch. 12.6).
13.1.1 Standard mode
This is the default setting. Use this mode unless stability or response has to be in-
creased or reduced.
Repeatedly press the [CAL] key until "Stnd".
Setting control
display
To set standard mode, press the [TARE] key.
Setting of this mode can be confirmed only in settings check dis-
(when selecting
play (see ch. 12.6).
standard mode)