17 Instant help
In case of an error in the program process, briefly turn off the balance and discon-
nect from power supply. The weighing process must then be restarted from the be-
ginning.
Error code table:
Error code display
CAL E2
CAL E3
CAL E4
CHE X (X is a numeral)
(when the display stops
here)
Err 0X (X is a numeral)
Err 20
Err 24
Possible causes of errors:
When
Symptom
Before
Nothing appears in the
meas-
display.
urement
During
The display fluctuates.
meas-
The standstill display
urement
does not appear fast
enough.
The measured results
have poor repeatabil-
ity.
"CAL d" appears fre-
quently.
ABT-NM-BA-e-1622
Explication
High zero point shift dur-
ing adjustment
Great deviation of meas-
uring values in PCAL.
Great deviation of meas-
uring values during ad-
justment
Internal malfunction
Internal malfunction
An improper value setting
was attempted.
Power voltage error
Possible causes
The AC adapter is not securely
connected. The power switchboard
of the room is turned off. The power
voltage is incorrect.
Vibrations or draft.
Remedy
Remove items from the
weighing pan.
Use correct adjusting
weight.
Please contact a service
representative.
Please contact a service
representative.
Enter the correct numbers
or decimal points.
Check the power voltage .
Remedy
Check power supply and
connect correctly.
Change the installation site.
Change stability and reac-
tion settings or standstill
display.
71