9 Servicing/repairs/cleaning
9.2
Repairs
9.3
Cleaning
EN-45
The load cell PR 6261 is designed to be as robust as possible for the required measuring
accuracy and is highly reliable. Should an electrical or mechanical defect nevertheless
occur, the load cell must be replaced.
Note:
See Chapter 9.1.2.
Load cell repair is not possible.
The weighing module is easy to clean. It can be spray-washed with water (see IP69
speciication).
For this purpose, spray the water jet from top to bottom and around the weighing
module.
NOTICE
Some cleaning agents may not be compatible with the weighing module material.
When using cleaning agents, ensure that their compatibility with the weighing
module material has been tested and approved (see Chapter 4.2).
Hygienic Weighing Module Novego®
Minebea Intec