5 Installation
5.1.3
Preparing vessel foot
EN-21
CAUTION
The pendulum base with tilt correction weighs approx. 4.2 kg.
Beware of any of a situation with light, reversible injuries as a result.
Always work with appropriate protective equipment.
1.
Position template (1) on the ground.
2. Remove the protective ilm of the two adhesive strips from the template.
3. Hold the lower plate (5) over the template (1) so that position 0° on the 0°‑marking of
the template and position 3° on the 3°‑marking of the template show up.
4. Place the lower plate without pendulum base onto the template and press.
The lower plate (5) is now ixed on the template.
5. Place the spirit level next to the marked ield (2) of the template (1).
6. Turn the template until the level reaches the marked target orientation.
7. Place the pendulum base (3) using the pin (in the upper plate) onto the lower plate so
that the 0°‑marking of the upper plate (4) shows up on the 0°‑marking of the lower
plate.
8. Position the spirit level on the center on any hole of the pendulum base.
9. Rotate the template counterclockwise, in small steps, and the upper plate (4) of the
pendulum base to the same extent in the clockwise direction until the bubble is
centered.
10. Position the drill holes (6) with a center mark onto the template.
11. Mark the upper plate in position 0° of the lower plate with a felt-tip pen.
12. Fold the template at the fold lines (7).
13. Mark the positions 0° and 3° of the lower plate using the guide lines (8) on the back of
the template with a felt-tip pen on the ground.
14. Remove template with pendulum base (3) from the ground.
15. Put holes for the screw anchors.
16. Then insert the threaded rods.
NOTICE
No leak-tightness of the supporting/sealing surface present between load cell and
weighing object.
It is essential to use adapter plate PR 6061/00S to prevent any negative efect on
the measuring properties.
If you do not use the adapter plate (see Chapter 12.3), the surface supporting the load cell
must fulill the following requirements:
-
Material: stainless steel
-
Flatness: 0.05 mm
-
The surface must not bend under the expected load (i.e.the deined latness may not
be exceeded). If the latness exceeds the set limit, the bending forces will transfer to
the mounted load cell and negatively afect its measuring properties. Especially its