5.3.2

Mounting the weighing module

Minebea Intec
Note:
The following describes how to mount the load cell with the adapter plate PR 6061/00S
(see Chapter 12.3).
The operations must be performed at all supporting points (e.g. vessel foot) of the
weighing object (e.g. vessel).
1.
Remove the O-ring (2) from the packaging of the load cell.
NOTICE
The restoring force is not suitable for the load cell.
Use only the O-ring supplied with the load cell.
2. Remove cap (1) of the pendulum base (3).
3. Insert the O-Ring (2) into the pendulum base and re-attach the cap.
Note:
The cap (1) lies loose on the pendulum base (3).
5 Installation
EN-24