5 Installation
5.3.2.5
Installation steps for ground installation with tilt correction
EN-33
1.
Slowly lower the weighing object (e.g. vessel) including the attached weighing
modules towards the bottom (21). While doing so, align the drill holes of the
pendulum bases (3) over the threaded bars (25).
2. Insert the O-ring (20) into the groove of the pendulum base (3).
3. Tighten pendulum base (3) using nut (24). Always comply with the strength classes
and tightening torques of the connector elements (see Chapter 5.2).
NOTICE
Never vertically straighten up a tilted pendulum base by pulling on the screws.
It is essential to observe a tolerance of ≤1 mm (see Chapter 5.1.2).
1.
Place the lower plate (20) using the 0°/3°‑marking onto the bottom (see
Chapter
5.1.2.2
operating step 11).