Unpacking TegraPol
Placing TegraPol
Supplying Water
Connection to Water Mains
Connection to Water Outlet
6
TegraPol-11/ -15
TegraForce-1
Instruction Manual
Detach TegraPol from the bottom of the packing case by removing
the four screws from below.
Place TegraPol on a steady table with an adequate working height.
The machine must be close to the power supply, water mains and
water outlet facilities.
Water for wet grinding may be supplied from the water mains or a
Recirculation Cooling Unit (instructions on the next page).
The cold water supply must have a head pressure in the
range 1 – 10 bar (14.5 – 145 psi).
Mount the 90° end of the inlet hose onto the water inlet tube on
the back of TegraPol (see
Insert the filter gasket in the coupling nut with the flat side
against the pressure hose.
Tighten the coupling nut completely.
Mount the straight end of the inlet hose on the water mains tap
for cold water:
Mount the reduction ring with gasket on the water mains tap,
if necessary.
Introduce the gasket and tighten the coupling nut
completely.
Mount the outlet hose onto the water outlet tube (Lubricate with
grease or soap to facilitate insertion). Use a hose clamp for
fastening.
If there is only limited space behind the TegraPol, use the pipe
bend and attach it directly to the outlet tube on the back of the
TegraPol. Then attach the drain hose to the other end of this
bend.
Lead the other end of the drain hose to the water outlet. Arrange
the tube so that it slopes downward towards the drain throughout
its length. Shorten the hose, if necessary.
Make sure that the drain hose slopes downward towards the drain
throughout its length and avoid sharp bends in the drain hose.
IMPORTANT
Getting Acquainted with
REMEMBER
TegraPol):