Centrifugation of materials or mixtures of materials with a density higher than 1.2 kg/dm

When centrifuging with maxim revolutions per minute the density of the materials or the material mixtures may not
3
exceed 1.2 kg/dm
.
The speed must be reduced for materials or mixtures of materials with a higher density.
The permissible speed can be calculated using the following formula:
Reduced
speed
(n
)
red
e.g.: maximum speed RPM 4000, density 1.6 kg/dm
1.2
kg/dm³
n
=
x
4000
red
1.6
kg/dm³
In the exceptional case that the maximum loading indicated on the hanger is exceeded, the speed must also be
reduced.
The permissible speed can be calculated using the following formula:
Reduced
speed
(n
)
red
e.g.: maximum speed RPM 4000, maximum load 300 g, actual load 350 g
300
g
n
=
x
4000
RPM
red
350
g
If in doubt you should obtain clarification from the manufacturer.
34

Rotor Identification

After every start of a centrifugation run the rotor utilised is identified.
After changing a rotor, the engine switches off and the rotorcode (rotor), the maximum rotor speed of the rotor
(Nmax) and the preset centrifuging radius (R) of the rotor being used are displayed, e.g. rotor 4 Nmax= 4500
R=184 mm .
Press the
key to open the lid, or press the
OPEN / STOP
with refrigeration, precooling of the rotor can also be started by pressing the
If, following a rotor change, the maximum speed of the rotor is less than the set speed, the speed is limited to
the maximum speed of the rotor. In this case, the program place number is displayed in parentheses ().
35

Emergency release

During a power failure the lid cannot be unlocked by motor. An emergency release has to be executed by hand.
For emergency release disconnect the centrifuge from the mains.
Open the lid only during rotor standstill.
See figure on page 2.
Switch off the mains switch (switch position "0").
Look through the window in the lid to be sure that the rotor has come to a standstill.
Insert the hexagonal wrench key into the bore hole (Fig. 1, A) and carefully rotate by half a turn in clockwise
direction until the lid can be opened.
Pull the hexagon socket head wrench out of the drilling again.
If the left LED in the
OPEN / STOP
button so that the motor-driven lid lock goes into the basic position (opened) again.
52/147
1.2
=
Greater
density
[kg/dm³]
3
RPM
=
3464
RPM
maximum
load
[g]
=
x
maximum
actual
load
[g]
=
3703
RPM
button flashes after the centrifuge is switched on again, press the
x
maximum
speed
[RPM]
speed
[RPM]
key to start the centrifugation run. For centrifuges
START
key.
3
OPEN / STOP