Les langues disponibles

Message / fault
SENSOR-ERROR
SENSOR-ERROR
91 - 93
NO ROTOR OR
ROTORCODE
ERROR
N > ROTOR MAX
N > ROTOR MAX
in Prog : e.g. 3
Runtime 00:00
in Prog : e.g. 3
Empty Program
FC INIT ERROR
FC VERSION ERROR
FATAL EEPROM
1 - 5
ERROR
38

Switch on automatic circuit breaker

(only for types 5650-01, 5660-01 and 5660-51)
Switch off the mains switch and separate the centrifuge from the mains!
The automatic circuit breaker is located in the lower right side of the front panel.
Push the plastic pin (A) of the automatic circuit breaker upwards.
Reconnect the centrifuge to the mains supply.
39

Acceptance of the centrifuges for repair

If the centrifuge is returned to the manufacturer for repair, it must be decontaminated and cleaned to protect persons,
environment and material.
We reserve the right to accept contaminated centrifuges.
Costs incurred for cleaning and disinfection are to be charged to the customer.
We ask for your understanding in this matter.
40

Disposal

Before disposal, the device must be decontaminated and cleaned to protect people, the environment and property.
When you are disposing of the device, the respective statutory rules must be observed.
Pursuant to guideline 2002/96/EC (WEEE), all devices supplied after August 13, 2005 may not be disposed as part of
domestic waste. The device belongs to group 8 (medical devices) and is categorized in the business-to-business
field.
The icon of the crossed-out trash can shows that the device may not be disposed as part of domestic
waste.
The waste disposal guidelines of the individual EC countries might vary. If necessary, contact your
supplier.
Cause
90
Error / defect electronics
Error / defect unbalance sensor
---
No rotor installed.
Faulty speedometer.
---
Speed in the selected program greater than
the maximum speed of the rotor.
There is a program in the displayed
program location whose speed is higher
than the maximum speed of the rotor.
There is a continuous operation program in
the displayed program location.
There is no program linkage saved on the
displayed program location.
Error / defect electronics
Error / defect electronics
Error / defect electronics
− Perform a MAINS RESET.
− Open lid.
− Install rotor.
− Check the set speed.
Reduce the set speed
− Replace the continuous operation
program in the program linkage
with a program with preset time.
− Call up a program linkage.
− Perform a MAINS RESET.