Installation

The digital Plug and Measure feature allows the user 
to measure CO
The sensor is recognized automatically by the trans-
mitter and all relevant data is sent to the transmitter 
for automatic configuration. The CO
displayed immediately.
The sensor is factory calibrated and the data is 
stored in the sensor. Calibration data therefore does 
not need to be entered by the user (to achieve best 
performance at given temperature we recommend to 
perform a process calibration).
4.1

Mounting the sensor

a
Important! Remove the protection cap before 
  m ount  i ng the sensor. 
Do not remove the MembraCap (membrane cap) 
that is screwed onto the sensor. If this is done the 
sensor needs recalibration.

4.1.1 Mounting and positioning the sensor

a
Attention! Do not turn the sensor counter-clockwise
when installed in a 28 mm housing. The Membra-
Cap might be loosened.
The sensor is available in three versions and can 
be fitted without a housing to the following process 
connections / armatures:
–  Sanitary DN40 VARINLINE
–  2" Tri-Clamp
–  28 mm M42 weld on sockets
–  28 mm M42 VARINLINE
The sensor must be installed in a way that the 
MembraCap is in contact with the process fluid to be 
measured.
–  Make sure the sensor is mounted from the side 
or from the top
–  Avoid mounting the sensor from the bottom
–  Mount the sensor in a flow ascending position
–  Avoid a position near to pump discharges 
Flow requirement:
For the sensor's proper functioning the flow in the 
pipe needs to achieve a min. flow of 0.5 m/s

4.1.2 Removing the sensor and storage

–  To remove the sensor from the process the pipe 
needs to be drained. Avoid counter-clockwise 
rotation in case where the sensor is installed on 
a 28 mm socket.
© 06 /14 Mettler-Toledo AG   
Printed in Switzerland 
 Sensor  
2
 levels immediately after installation.
2
 Sanitary Connector
®
 level is then 
2
®
 Type N Connection
 stationary adapters
®
13
InPro 5500 i 
30 035 328

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Co2 sensor

Table des Matières