Ism Diagnostics - Mettler Toledo InPro 5500i Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

20 
Example:
Sensor verification using a transmitter and a
calibration gas:
–  Press Cal button "Calibrate Sensor" and confirm.
–  Chose "1-point CO
–  Start calibration when sensor is in gas.
–  Enter the partial pressure of the calibration gas.
–  The transmitter will now show the slope  
S = **mV and the baseline values 
BL = **mV and asks for "Save adjust".
–  If you want the sensor to measure based on the 
new calibration data choose "Save adjust".
–  If you want only to save the calibration data 
chosse "Save calibrate" (the sensor in this case 
still measures with the previous adjusted cali-
brtion data).
–  for verification purposes with a calibration gas 
the slope values can be compared with the 
previous calibration values. The previous 3 cali-
brations are saved in the ISM-Info menu ( Info 
Menu: Cal Data)
h
Note: The sensor now still uses the previous calibra-
tion data for the measurements. The sensor will only 
measure with the new calibration data by using the 
command "save adjust". "Safe calibrate" stores the 
calibration data for verification tracking purposes 
only.
h
Note: If the notice "cal out of limit" appears, please 
check the purge gas supply and calibration gas sup-
ply. If you still get the notiice, replace the Membra-
Cap and perform a new calibration.

6.1.4 ISM diagnostics

Dynamic Lifetime Indicator: DLI
The DLI provides information about the remaining 
lifetime of the MembraCap. As long as the DLI is 
above zero days the system is within the specified 
accuracy after a calibration. If the DLI is zero the 
MembraCap shall be replaced. Contributing factors 
for wear of the MembraCap are:
–  temperature during measurement
–  number of CIP cycles
–  number of SIP cycles
DLI calculation: Using the above parameters actual 
sensor stress is calculated. The accumulated sensor 
load divided by the elapsed time is the basis of the 
calculation of the remaining lifetime.
InPro 5500 i 
30 035 328 
InPro 5500 i CO
" calibration and confirm.
2
© 06 /14 Mettler-Toledo AG
Printed in Switzerland
 Sensor
2

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Co2 sensor

Table des Matières