Calibration; Purpose Of Calibration; Single Point Calibration In-Line, Off-Line - Mettler Toledo InPro 5500i Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

InPro 5500 i CO
5.3

Calibration

5.3.1 Purpose of calibration

Information about the calibration, you find also in 
the manual of the transmitter. 
Calibration should be performed at least after each 
change of the MembraCap.
Since the measurement of the thermal conductivity is 
dependent on the diffusional properties of the used 
MembraCap, a calibration after exchange is manda-
tory to regain accurate CO

5.3.3 Single point calibration in-line, off-line

The sensor is designed to provide accurate CO
measurement with a simple slope calibration using 
a pure CO
partial pressure of the CO
The calibration can be performed in-line or by dis-
mounting the sensor and performing the calibration 
in the workshop.
METTLER TOLEDO offers a calbox (P/N: 52 300 400) 
as a support for the sensor for workshop calibration. 
The CalBox has a 6 mm hose connection to connect 
the calibration gas. 
h
Note: If the calibration is done in the workshop, 
make sure a purge air source is used.
h
Note: The air supply (purge air) needs to be active 
at all times while performing a calibration.
Calibration procedure:
–  Start the calibration by pressing the cal button on 
the transmitter and chosing "1-point calibration"
–  Process pipe needs to be filled with the calibra-
tion gas 
–  After exposing the sensor to the calibration gas 
confirm on the transmitter that the "Sensor is in 
Gas"
–  Enter the known CO
bration menu in mbar or hPa and continue 
–  Per default the sensor starts calibration when 
CO
–  For user convenience the Auto-Cal mode can 
be changed to manual cal mode. In this case 
the calibration can be started as soon as the 
sensor delivers a measured value (approx. every 
20 sec.), provided a stable reading is given
–  Once the calibration is done, save the calibration 
onto the sensor by selecting "Save Adjust"
–  A message "calibration successful" appears and 
the user is asked to "Re-install sensor" 
–  Now the sensor is recalibrated and ready for use
© 06 /14 Mettler-Toledo AG   
Printed in Switzerland 
 Sensor  
2
 gas or using a gas mixture where the 
2
 measurement is stabilized (Auto-CAL Mode)
2
 measurements.
2
 content is known.
2
 partial pressure in the cali-
2
17
2
InPro 5500 i 
30 035 328