Operation

5.1

Start-up

Each sensor is supplied ready to use. Before using 
remove the protecting cap. 
5.2

Configuration

5.2.1 Sensor detection

Before installing the CO
manual for the transmitter and configure the 
transmitter for automatic sensor detection. The 
M800 transmitter does not need any pre-configura-
tion. 
5.2.2 CO
solubility and temp. factor
2
The sensor is delivered factory calibrated and is set 
up to measure in beer as the default.
The sensor offers a choice of CO
surement in beer, cola or water. For other beverages 
the user has the possibility to enter individual values 
for CO
Default values for measurement in beer (Valid for 
temperatures – 5 ... 50 °C): 
CO
 solubility (A):  1.420 g / L
2
Temp. factor (B):  2485 
Values for pure water: 
CO
 solubility (A):  1.471 g / L
2
Temp. factor (B):  2491
Values for cola: 
CO
 solubility (A):  1.345 g / L
2
Temp. factor (B):  2370
For beverages where the user knows the exact CO
solubility and the temperature factor the values can 
be changed individually. If the user desires to evau-
ate the solubility and temperature factors they can 
be evaluated with the following formulas. 
H
 = A  *  exp (B  *  (1 / T – 1 / 298.15)
CO2
c
 = H
CO2
H
 : 
CO2
   
A:  
B:  
c
: 
CO2
InPro 5500 i 
30 035 328 
 solubility and temperature factor.
2
  *  p
CO2
CO2
Calculated CO
constant) at measured process temp.
Solubility of CO
Temperature factor (valid for – 5 – 50 °C) 
Calculated CO
InPro 5500 i CO
 sensor, please refer to the 
2
 solubility for mea-
2
 solubility (Henry 
2
 (g / L at 25 °C)
2
 concentration
2