Note: If the message "Cal out of Limit" appears 
please check the gas supplies (calibration gas and 
purge gas) and / or if the MembraCap needs to be 
exchanged. In this case please abort the calibration.
h
Note: If only a verification of the calibration is need-
ed, choose "safe calibrate" in the transmitter menu. 
There is also the choice for aborting the calibration 
"save abort".
h
Note: When the MembraCap is exchanged the 
sensor needs to be operated at least 15 min. before 
performing a calibration.
h
Note: Be aware that when calibration is done with 
CO
 reference gas the sensor will show sligthly dif-
2
ferent CO
the sensor's electronic use a compensation to give 
accurate measurement in liquids. For calibration 
verification purposes with reference gas please fol-
low the steps in chapter 6, comparing the new and 
old slope values. 

5.3.4 In-line process single-point calibration

If the user requires that the sensor measures the 
same values as with established off-line reference 
CO
 measuring methods, the InPro 5500 i offers a 
2
single-point process calibration option.
Calibration procedure:
–  Start calibration by pressing the cal button on the 
transmitter and select CO
Type = Process
–  Take a sample from the process fluid at the same 
time from which the CO
mined by an established reference method and 
confirm capture
–  After analyzing the CO
transmitter and start cal again and enter the 
reference value in mg / L, V / V, mbar or hPa
–  Once the calibration is done, save the calibration 
to the sensor by confirming "Save Adjust"
–  A message "calibration successful" appears 
–  Now the sensor is process calibrated and ready 
for use
h
Note: If message "Cal out of Limit" appears please 
check your reference measurement or if the mem-
brane cap needs to be exchanged. In this case 
please abort the calibration.
InPro 5500 i 
30 035 328 
 values in gas. The reason for this is that 
2
InPro 5500 i CO
 calibration 
2
 value has to be deter-
2
 value go back to the 
2