InPro 5500 i CO
1.  Dry sensor carefully with a soft tissue
2.  Unscrew the MembraCap sleeve from the sensor 
shaft and carefully pull it off the sensor (Fig. 4).
3.  Place the new MembraCap on top of the inner 
body of the sensor shaft (Fig. 5).
4.  Carefully slip the MembraCap over the sensor 
and screw it down as tightly as possible by hand 
(do not use a tool) [Fig. 6 to 7]. 
5.  After each exchange of the MembraCap, the DLI  
must be reset manually using the transmitter.
6.  After each exchange of the MembraCap, the sen-
sor has to be recalibrated by a 1-point calibration 
using a CO
tion in measuring solution. Make sure to reset 
DLI before performing a calibration. The sensor 
will automatically prompt for calibration.
Fig 4: Unscrewing MembraCap counterclockwise
Fig. 5: Placing the new MembraCap on top of the 
sensor