When changing the MembraCap, please observe the 
following instructions:
a
Attention! Make sure that this maintenance step is
carried out in a clean environment.
a
Attention! Make sure you hold the sensor upside 
down (Fig. 1, 2 and 3). Otherwise the diffusor might 
fall out and the thermal conductivity chip will be 
exposed. If the diffusor falls out clean it gently with 
alcohol and place it back on top of the sensor.
Fig. 1: Hold sensor in upside down position during 
maintenance e.g. MembraCap change
Fig. 2: Sensor view with removed diffusor
Fig. 3: Sensor with removed MembraCap
InPro 5500 i 
30 035 328 
InPro 5500 i CO
© 06 /14 Mettler-Toledo AG
Printed in Switzerland