InPro 6800 Series O
4
Installation
4.1
Mounting the sensor
Important! Remove the green protective cap before
mount ing the sensor.
Mounting the sensor in a housing
Please refer to the instruction manual of your housing
explaining on how to mount the sensor in place.
Mounting the sensor directly on a pipe or a vessel
The sensor can be mounted vertically or in an inclined
position. In the case of inclined mounting, the angle of
the housing must be equal to or greater than 15° above
the horizontal.
Caution! Installation outside the admissible range of
mounting positions is not allowed, otherwise correct
operation of the sensors / electrodes may be impaired.
The 12 mm sensors can be mounted directly through
a socket with inside thread Pg 13.5 and securely tight-
ened via the Pg 13.5 threaded sleeve.
The 25 mm sensors can be mounted directly through
a standard weld-in socket or the safety weld-in sockket
from METTLER TOLEDO.
© 12 / 2018 Mettler-Toledo GmbH
Printed in Switzerland
Sensor 12/25 mm
2
Admissible mounting positions
Zulässige Einbaulagen
Positionements de montage admis
Admissible mounting position
Zulässige Einbaulage
Positionement de montage admis
Inadmissible (except for optical sensors)
Unzulässig (ausser für
optische Sensoren)
Ne pas admissible (sauf pour
sondes optiques)
19
InPro 6800
52 200 953 B