How To Use By Front Panel - Agilent Technologies Turbo-V 301-AG Notice De Mode D'emploi

Table des Matières

Publicité

How to Use by Front Panel

Figure 29
The front panel START/STOP function is available only if the controller is set to
NOTE
"FRONT mode". If it is set to "REMOTE mode", when the key is pressed an error
message is shown indicating "COMMAND LOCKED IN REMOTE", if it is set to
"SERIAL" mode, the "COMMAND LOCKED IN SERIAL MODE" error message is
displayed. The "Reset Fail" function is always available regardless of controller
mode.
The front panel LOW SPEED function is available only if the controller is set to
NOTE
"FRONT"; otherwise an error message is returned. If it is set to "REMOTE"
mode, the COMMAND LOCKED IN REMOTE error message is displayed while if
it is in "SERIAL" mode the "COMMAND LOCKED IN SERIAL MODE" error
message is returned.
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
Start Stop and Reset Fail button
Press once for the START command and again for the STOP
command. In the case of controller's failure, press once to reset
the fail condition.
Low Speed
Press once to activate the LOW SPEED command; press again to
de-activate.

How to Use by Front Panel

LS
R1
AG
R2
VACUUM TECHNOLOGIES
COUNTERS
MEASURES
increase
menu
decrease
previous
LOW SPEED
START/STOP
Turbo-V 301-AG
Technical Information
next
245/294
17

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969899196989929698993

Table des Matières