Technical Information
Profibus Option
7
2
mA
9
2
W
11
2
W
13
2
Hz
15
2
Hz
17
2
-
19
2
min
21
2
h
23
4
mBar/Torr/
Pa
Diagnostic Management
In addition to the ProfiBus six byte standard diagnostic, the interface
provides both user diagnostic and static diagnostics functions
The standard diagnostic management is fully compliant with the
ProfiBus specification.
During the start-up phase after power on, the first byte in the user
diagnostic area, can assume values different from zero.
This byte has the following meaning:
280/294
0-10000
Current
Motor current consumption
0-200
Power
Motor power adsorption
0-200
Power Limit Applied
Maximum allowable power
0-1500
Driving Frequency
Read the rotational speed of the pump
0-1500
Target Frequency
Read the target rotational speed
0-65536
Cycle Number
Number of cycles (start and stop) done by the pump
0-65536
Last Cycle Time
How long the last cycle lasted
0-65536
Pump Life
Total pump running time
-
Pressure Reading
Reports the pressure reading in the "pressure unit"
selected by Par. Buff., Offset-20, B
Bit 0 = 1: controller not responds (RS485 link broken)
Bit 1 = 1: controller not identified (not compatible with the
ProfiBus gateway)
Bit 2 = 1: first loop of requests not ended (values in the input data
are not fully coherent yet)
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)