Parameter Buffer - Agilent Technologies Turbo-V 301-AG Notice De Mode D'emploi

Table des Matières

Publicité

17
Technical Information
Profibus Option

Parameter Buffer

Tab. 18 Parameter buffer
OFFSET
SIZE
UNIT RES
BYTE
BYTE
0
1
-
1
2
mA/Hz/sec
3
2
Sec
5
2
mA/Hz/sec
7
2
Sec
9
2
0.2sec/bit
11
2
0.2sec/bit
13
1
%
14
1
%
274/294
RANGE
DESCRIPTION
0-99
Gauge gas load type. The value is divided by 10
(entering 12 the controller will apply 1.2)
Enter here the correction factor for gas (necessary
only if gauge is connected).
0-10000
Setpoint 1 value. Enter here the threshold value for
the R1 output, valid only if B
Buffer Offset-21 is equal to 0,1,2 or 3.
0-30000
Setpoint 1 Mask Time. Set here the time, measured
from the start pump event, while the R1 output is
masked (not activated)
0-10000
Setpoint 2 value. Enter here the threshold value for
the R2 output, valid only if B
Buffer Offset-21 is equal to 0,1,2 or 3.
0-30000
Setpoint 2 Mask Time. Set here the time, measured
from the start pump event, while the R2 output is
masked (not activated)
0-30000
Vent Valve Opening Delay.
Set the delay time between the stop command and
vent opening. Valid only in "purge/vent auto mode"
(Parameter Buffer, Offset 20, B
(the valve will never open).
0-30000
Vent Valve Open Time. Set how long the vent will
remain open, then both purge and vent will be
closed. Valid only in "purge/vent auto mode"
(Parameter Buffer, Offset 20, B
(the valve remain open till next start command).
1-99
Setpoint 1 Hysteresis. Set the Setpoint1 hysteresis
around the nominal value
1-99
Setpoint 2 Hysteresis. Set the Setpoint2 hysteresis
around the nominal value
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
-B
of Parameter
0
2
-B
of Parameter
3
5
= 0). 0 = infinite
0
= 0). 0 = infinite
0
WIN
SERIAL
161
102
103
172
173
126
147
105
175

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969899196989929698993

Table des Matières