Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
0-99
Setpoint 1 Pressure Value Mantissa.
Set the mantissa of the exponential value of the R1
setpoint (format X.XE-XX) when it is used for
pressure monitoring (Parameter Buffer, Offset 21,
B
= 4). The value set here is divided by 10, that
0-2
means 11 = 1.1
-10 - +10
Setpoint 1 Pressure Value Exponent.
Set the exponent of the exponential value of the R1
setpoint (format X.XE-XX) when it is used for
pressure monitoring (Parameter Buffer, Offset 21,
B
= 4).
0-2
0-99
Setpoint 2 Pressure Value Mantissa.
Set the mantissa of the exponential value of the R2
setpoint (format X.XE-XX) when it is used for
pressure monitoring (Parameter Buffer, Offset 21,
B
= 4). The value set here is divided by 10, that
3-5
means 11 = 1.1
-10 - +10
Setpoint 2 Pressure Value Exponent.
Set the exponent of the exponential value of the R2
setpoint (format X.XE-XX) when it is used for
pressure monitoring (Parameter Buffer, Offset 21,
B
= 4).
3-5
B
= Stop Speed Reading.
0
This function measures the pump speed after a stop
command.
0 = Off
1 = On
B
= Active Stop.
1
This function brakes the pump after a stop
command, by using the motor as a generator.
0 = Off
1 = On
B
- Reserved
2
B
= Start/Stop/Interlock Mode.
3-4
Change the logic of Start/Stop and Interlock inputs.
Ask to Agilent personnel for this option
Technical Information
Profibus Option
275/294