Letter Protocol Description
The default protocol is "Window"
Communication format:
Communication protocol:
Host = Master
Turbo-V 301 Controller = Slave
The communication is performed in the following way:
Host
Request is an ASCII character identifying the action that must be
performed by the controller or the requested information.
Allowed ASCII characters:
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
8 data bit
no parity
1 stop bit
The baud rate is programmable via front panel from 600 to 9600
baud. The controller is factory-set for 9600 baud operation.
"A" = START
"B" = STOP
"C" = Low Speed ON
"D" = Low Speed OFF
"E" = Request for operational parameters
"F" = Pump times zeroing
Technical Information
Letter Protocol Description
Turbo-V 301 Controller
17
229/294