Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
Pressure unit of measure
Stop speed reading
Activates / deactivates the pump speed
reading after Stop command
R2 Set Point Type
R2 Set Point Value (Hz, W, s)
R2 Set Point Mask (sec)
R2 Set Point Signal Activation Type
R2 Set Point hysteresis (in % of R2 Valve)
R2 Set Point Pressure Threshold
Valid if win 171 = 4
Format X.X EsX.X
Where: X= 0 to 9
s = + or -
Start Output Mode
Pump current in mA dc
Pump voltage in Vdc
Pump power in W (pump current x pump
voltage duty cycle)
Driving frequency in Hz
Pump temperature in °C
Pump status
Error code
Controller Heatsink Temperature (°C)
Technical Information
Examples
0 = mBar
1 = Pa
2 = Torr
0 = disable
1 = enable
0 = Freq
1 = Power
2 = Time
3 = Normal
4 = Pressure
Ø = high level active
1 = low level active
0 = Starting (Output ON
only with pump
Status = Starting) DEFAULT
1 = running (Output ON
when the pump
is running)
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see the
following figure