17
Technical Information
Letter Protocol Description
CRC corresponds to the sum (with inverted sign) of all the preceding
bytes.
e.g., the START command "A" in ASCII code = 41 inverted it will be:
FF + 1 - 41 = BF.
Tab. 6
CHARACTER
"A"
CRC
"B"
CRC
"C"
CRC
"D"
CRC
"E"
CRC
"F"
CRC
"G"
CRC
"H"
"I "
CRC
"J"
CRC
230/294
"G" = Parameters reading
"H" = Parameters writing
"I" = Request for operating status
"J" = Request for numerical reading
"K" = Request for counters reading.
ASCII
41h
-
42h
-
43h
-
44h
-
45h
-
46h
-
47h
-
48h
49h
-
4Ah
-
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
CONVERTED ASCII FOR CRC CALCULATION
-
BFh
-
BEh
-
BDh
-
BCh
-
BBh
-
BAh
-
89h
-
-
B7h
-
B6h

Publicité