Press the CYCLE NUMBER twice and the display
shows:
X X X X C Y C L E
P U M P
L I
where:
CYCLE = are the cycles performed (range 0 to 9999)
m = is the elapsed time related to the cycle number
displayed (range 0 to 99999 minutes)
PUMP LIFE = is the total operation time of the pump
(range 0 to 99999 hours).
X X X X X m
F E
X X X X X h

FRONT / REMOTE Selection

Press CYCLE NUMBER and PUMP CURRENT
pushbuttons together for at least 2 seconds and the
processor enters in a routine where it is possible to
program the controller.
In this routine, the CYCLE pushbutton is used for
choosing/changing the value or condition; the
PUMP CURRENT pushbutton is used to enter and
confirm the value. At any time it is possible to exit
this routine by pressing the CYCLE and PUMP
CURRENT pushbuttons at the same time for at
least 2 seconds.
The display shows:
The display shows:
S O F T
S T A R T
S E L E C T I
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump
to rump-up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is
deselected and the rump-up of the pump will be done
within 360 seconds.
The controller is factory set to YES.
The Soft Start mode may be deselected/selected only
when the pump is stopped.
Press CYCLE NUMBER to select YES or NO
S O F T
S T A R T
S E L E C T I
84
TECHNICAL INFORMATION
M O D E
O N :
X X X
NOTE
M O D E
O N :
Y E S
87-900-906-01 (E)