17
Technical Information
Examples
110
R/W
L
111
R/W
L
117
R/W
N
120
R/W
N
122
R/W
L
123
Reserved to Agilent service
124
125
R/W
L
126
R/W
N
130
Reserved to Agilent service
143
R/W
N
144
R/W
L
147
R/W
N
155
R
N
157
R/W
N
162
R/W
A
242/294
Interlock type (default = 0)
Analog output type: output voltage
signal proportional to frequency, power,
temperature or pressure
Low Speed frequency (Hz)
Rotational frequency setting (Hz)
Set vent valve on/off (on = closed)
Set the vent valve operation
Vent valve opening delay (expressed in 0.2
sec)
External Fan Configuration
External Fan Activation
Vent open time
See "vent connector" paragraph
Power limit applied
Read the maximum power deliverable
to the pump
Gas load type.
Select the gas load to the pump
R1 Set Point Pressure Threshold
Valid if win 101 = 4
Format X.X EsXX
Where X = 0 to 9 s = + or -
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 1)
2 = Pump temperature
3 = Pressure
700 to rotational frequency
setting (win 120)
(default = 700)
Low Speed frequency
(win 117) to 963
(default = 963)
On = 1
Off = 0 (default = 1)
Automatic = 0 (see note 1.)
On command = 1 (see note
2.)
0 to 65535 (corresponding to
0 to 13107 sec)
0 = ON
1 = automatic
2 = serial
(default = 2)
0 = OFF
1 = ON
(default = 0)
0 = infinite
1 bit = 0.2 sec
watt
Ø = N
2
1 = Av