17
Technical Information
Profibus Option
Input Buffer
Tab. 20 Input buffer
OFFSET
SIZE
UNIT RES
BYTE
BYTE
0
1
V
1
1
°C
2
1
°C
3
1
°C
4
1
-
278/294
RANGE
DESCRIPTION
0-100
Voltage
Voltage provided to the motor
0-100
Pump Bearing Temperature
Read the pump upper bearing temperature
If > 60 °C the controller goes into fail status and the
pump is stopped
0-100
Controller Power Supply Temperature
Read the controller power section temperature if >
75 °C the controller goes in fail status "controller
over-temperature"
0-100
Controller CPU Section Temperature
Read the controller CPU section temperature if
> 80 °C the controller goes in fail "controller
overtemperature"
0-6
B
– Status
0-3
0 = Stop
1= Waiting for interlock (interlock connections on
remote I/O connector missing)
2 = Ramp (the pump is increasing the speed)
3 = Auto tuning (the speed has been reduced
because the gas load is higher than the Power Limit
Applied)
4 = Braking (the pump is reducing the speed using
the motor)
5 = Normal operation (the pump is at target speed)
6 = Fail (see error code byte Input Buffer Byte-27)
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
WIN
SERIAL
201
204
216
211
205