Installation
WARNING!
The Turbo-V controller must be powered with 3-wire power cord (see
orderable parts table) and plug (internationally approved) for user's safety.
Use this power cord and plug in conjunction with a properly grounded power
socket to avoid electrical shock and to satisfy CE requirements. High voltage
developed in the controller can cause severe injury or death. Before servicing
the unit, disconnect the input power cable.
The Turbo-V controller can be used as a bench unit or a rack module, but it must
NOTE
be positioned so that free air can flow through the holes. Do not install or use
the controller in an environment exposed to atmospheric agents (rain, snow,
ice), dust, aggressive gases, or in explosive environments or those with a high
fire risk.
During operation, the following environmental conditions must be
respected:
See the appendix "Technical Information" for detailed Information
about the above mentioned and the other connections, and about the
options installation.
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
temperature: from +5 °C to +45 °C;
relative humidity: 0 – 95 % (without condensation).