Pins
Name
2-1
Programmable
Analog Output
9-1
Frequency
Analog Output
When the analog output is set to provide pressure signal, the voltage
level is proportional to pressure level and it is generated by control
unit elaborating the independent input by the active gauge.
If the Programmable Analog Output is related to the pressure read by
an external gauge, the controller doesn't replicate exactly the signal
coming from the gauge; the controller converts the gauge signal in the
range from 0 to 10V.
The conversion formula is the following:
V
out
Where:
Vg is the gauge output voltage;
Vout is the Analog Output voltage;
V
g_min
can be drawn from the gauge technical specifications).
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01 (I)
Description
Programmable analog output.
Can be related to : frequency,
power, pump temperature
(linear) or pressure read
by the external gauge.
The setting can be done by:
front panel or serial line
(win.111).
0V = 0Hz, 0W, 0°C
10V = 963Hz, 150W, 100°C
See following chart for
more details.
Analog output proportional to
frequency (linear).
0V = 0Hz
10V = 963Hz
V −V
g
g
_ min
=
×
−
V
V
g
_ max
g
_ min
V
÷
is the gauge voltage measuring range (these values
g_max
Electrical
Optically isolated
Rload, min = 10kohm
Ripple Vpp,max =
200mV
Precision = 5%
Resolution = 40mV
Pin-1negative
Pin-2 positive
Optically isolated
Rload, min = 10kohm
Ripple Vpp,max =
200mV
Precision = 1%
Resolution = 40mV
Pin-1negative
Pin-9 positive
10
Technical Information
Controller Connection
Note
Connect a 100kohm
resistor between the 2
pins before measuring
voltage
219/294