Annex 1: ELECTROMAGNETIC COMPATIBILITY
Somnibel Pro
is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment.
Emissions test
RF (Radiated) emissions
CISPR 11 (EN 55011)
RF (Conducted) emissions
CISPR 11 (EN 55011)
Harmonic emissions
EN-IEC 61000-3-2
Voltage fluctuations / Flicker emissions
EN-IEC 61000-3-3
EN 23
Guidance and manufacturer's declaration – electromagnetic emissions
Compliance
Group 1
Class B
Grupo 1
Clase B
Not applicable
Yes
somnibel Pro
Electromagnetic environment - Guidance
Somnibel Pro
uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
Somnibel Pro
works with batteries.
Applicable to the charger.
Somnibel Pro
works with batteries.
Power of the charger is lower than 75W.
Somnibel Pro
works with batteries.
Applicable to the charger.
534-80P-MUM_Rev.1.08