PNs 9698404, 9698405, 9698406, 9698419, 9698430, 9698431, 9698432, 9698433,
NOTE
are not equipped with the selector named "Pumping", the same function is
provided by the pushbutton Start/Stop on the TMP controller.
TPS-mobile-switching-off:
▪
▪
▪
TPS-mobile User Manual / 87-901-005-01 (F)
You can switch the pumps off by using the "pumping" selector
(moving it to position "0") – the TMP controller will remain ON
and if a gauge is connected to the system it will be possible to
continue reading the pressure even if pumps are off.
An alternative mode for switching the pumps off is the use of
Main Switch (move it to position "0") – The system will be
completely switched off and no reading or operation will be
possible.
If an emergency situation occurs is possible to switch the pumps
and the controller off, even disconnecting the mains cable
(emergency stop).
Technical Information
Appendix 1
393/400
17