Window Meanings
N.
Read/
Data
Write
Type
000
R/W
L
001
R/W
L
008
R/W
L
100
R/W
L
101
R/W
N
102
R/W
N
103
R/W
N
104
R/W
L
105
R/W
N
107
R/W
L
108
R/W
N
109
W
L
110
R/W
L
TPS-mobile User Manual / 87-901-005-01 (F)
Description
Start/Stop (in remote/ Front mode the
window is a read only)
Low Speed Activation
Remote (default) or Serial configuration
Soft Start (write only in Stop condition)
R1 Set Point type
R1 Set Point valve (expressed in Hz, W or s)
Set Point delay: time between the pump
start and the set point check (seconds)
Set Point signal activation type: the signal
can be "high level active" or "low level
active"
Set point hysteresis (in % of value)
Active Stop (write only in stop)
Baud rate
Pump life/ cycle time/ cycle number reset
Interlock type (default = 1)
Technical Information
Letter Protocol Description
Admitted Values
Start = 1
Stop = 0
No = 0
Yes = 1 (default = 0)
Remote = 1
Serial = 0 (default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Power
2 = Time
3 = Normal
4 =Pressure (default = 3)
(default = 900)
0 to 99999 (default = 0)
0 = high level active
1 = low level active
(default = 0)
0 to 100 (default = 2)
0 = NO
1 = YES
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
17
383/400