17
Technical Information
Electrical Connections
The communication port mating connector is supplied with the RS
232 PCB (AMP/Cannon or equivalent 15-pin "D" type male
connector).
For example, the Transmit data signal from controller (pin 2) must be
connected to the host computer's receive data line (pin 2) and vice
versa. Consult the host computer's instruction manual for its serial
port connections.
Agilent cannot guarantee compliance with FCC regulations for radiated
NOTE
emissions unless all external wiring is shielded, with the shield being
terminated to the metal shroud on the 0-subconnector. The cable should be
secured to the connector with screws.