163
R/W
N
167
R/W
A
171
R/W
N
172
R/W
N
173
R/W
N
174
R/W
L
175
R/W
N
176
R
A
200
R
N
201
R
N
202
R
N
203
R
N
204
R
N
205
R
N
206
R
N
211
R
N
TPS-mobile User Manual / 87-901-005-01 (F)
Pressure unit of measure
Stop speed reading
Activates / deactivates the pump speed
reading after Stop command
R2 Set Point Type
R2 Set Point Value (Hz, W, s)
R2 Set Point Mask (sec)
R2 Set Point Segnal Activation Type
R2 Set front Mysteresis (in % of R2 Valve)
R2 Set Point Pressure Valve
Valid in win 171 = 4
Format X.X E X.X
Where: X= 0 to 9
s = + or -
Pump current in mA dc
Pump voltage in Vdc
Pump power in W (pump current x pump
voltage duty cycle)
Driving frequency in Hz
Pump temperature in °C
Pump status
Error code
Controller Heatsink Temperature ( °C)
Technical Information
Letter Protocol Description
0 = mBar
1 = Pa
2 = Torr
0 = disable
1 = enable
0 = Freq
1 = Power
2 = Time
3 = Normal
4 = Pressure
Ø = high level active
1 = low level active
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see the
following figure
17
385/400