When connected to a PC, to enable logging:
 Press LOG/CLEAR. LCD displays "LOG ON METER" with ACCEPT
tag blinking.
 Press GLP/ACCEPT. Meter disconnects from the PC and the PC
icon is no longer displayed.
 To return to "EXPORT TO PC" mode, follow steps 2 and 3 above.
Exported data file details:
 The CSV file (comma separated values) may be opened with a
text editor or spreadsheet application.
 The CSV file encoding is Western Europe (ISO–8859–1).
 Field separator may be set as comma or semicolon.
See Separator Type (MW106) in SETUP OPTIONS section.
 Interval log files are named PHLOT###, where ### is the lot
number (e.g. PHLOT051).
 Manual log file is named PHLOTMAN and stability log file is
named PHLOTSTA.
USB Export All
1. With the meter on, insert a USB flash drive into the micro USB
port located on top of the meter. If the flash drive does not
have a micro USB connector, use an adapter.
2. Press RCL then RANGE/ to select the "EXPORT ALL" option.
3. Press GLP/ACCEPT to confirm.
LCD displays "EXPORTING" and the percentage counter,
followed by "DONE" when export is completed. Display returns
to the lot selection screen.
Note: The USB flash drive can be safely removed if the USB icon
is not displayed. Do not remove the USB drive during export.
LOGGING (MW106) 39