Guidelines and Manufacturer's Statement - Electromagnetic Compatibility
en
48
b
The field strength should be less than 3 V/m for the frequency range of 150 kHz
to 80 MHz.
12.5
Recommended Safety Distances Between Portable and Mobile
RF Telecommunications Devices and Aquilex Fluid Control Sys-
tem
The Aquilex Fluid Control System is intended for use in an electromagnetic envi-
ronment where HF interferences are controlled. The user/operator of the Aquilex
Fluid Control System can contribute to lowering electromagnetic emissions by
complying with the minimum distance between portable and mobile HF tele-
communications devices (transmitters) and the Aquilex Fluid Control System, de-
pending on the output power of the communication equipment listed below.
Rated output of
Safety distance based on the transmitting frequency [m]
the transmitter
150 kHz to
[W]
80 MHz
d = 1.2P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
The safety distance d in meters [m] for transmitters with a max. rated output not
listed in the table above can be calculated by applying the corresponding formula
in the respective column. P is the max. rated output of the transmitter in watts
[W] as specified by the transmitter manufacturer.
Note 1: The higher frequency range applies at 80 MHz and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
WARNING!
Portable HF communication equipment can affect the performance characteris-
tics of the Aquilex Fluid Control System. Such equipment must therefore comply
with a minimum distance of 30 cm (regardless of all calculations) from the Aqui-
lex Fluid Control System, its components and cables.
80 MHz to
800 MHz to
800 MHz
2.7 GHz
d = 1.2P
d = 2.3P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23