9.3
Determine the Software Version
The software version of the pump may be required for further analysis.
Follow these steps to determine the software version:
1. Turn device on and wait for the device self-test.
2. Press the Pause/Resume button for at least 2 seconds.
3. The software version of the pump is depicted in the "Deficit Limit" display and
consists of a 5-digit numerical sequence.
Enter the determined software version into the test log in Section 16.1.
Aquilex Fluid Control System
TM
High
Ser. Nr. WOM
/
100-240 V / 50-60 Hz
YYYY-MM
Potential
Equalization
(1)
Fig. 9-4
Low
H
I
G
H
Medium
only for
®
MyoSure
Annual Inspection
Vacuum pump exhaust ports
en
39