Télécharger Imprimer la page

Hettich EBA 270 Mode D'emploi page 66

Masquer les pouces Voir aussi pour EBA 270:

Publicité

Les langues disponibles

Les langues disponibles

Troubleshooting
1
Fig. 16: Emergency release
1
Hole
9.4
Replacing the mains input fuse
1
Fig. 17: Mains input fuse
1
Fuse holder
2
Snap lock
30 / 32
WARNING
Danger of cutting and crushing due to moving rotor.
The plastic unlocking pin supplied may only be used for
emergency device release. The release pin must be stored
in such a way that it is protected against unauthorised
access.
Personnel:
Trained user
1.
Look through the window in the lid to ensure that the rotor is sta-
tionary.
2.
Insert the supplied plastic release pin horizontally into the hole (
Push it in until the lid can be opened when the pin is pressed down.
3.
Open the lid.
Risk of electric shock due to maintenance and servicing
work on live device.
Personnel:
Trained user
The mains fuses are located next to the mains switch.
2
The mains switch is in switch position
1.
Disconnect the mains cable from the device plug.
2.
Press the snap locks (
out.
3.
Replace the defective mains input fuses.
2
Only use fuses with the nominal value specified for the type: see the
table below.
4.
Push in the fuse holder (
5.
Reconnect the device to the mains.
Model
EBA 270
Rev.: 12 / 10.2023
Do not open the lid until the rotor has stopped.
WARNING
Disconnect the device from the mains before carrying
out repairs and maintenance.
[O]
2  ) against the fuse holder ( 1  ) and pull them
1  ) until the snap lock engages.
Type
Fuse
2300
T 1.6 AH/250 V
1  ).
Order no.
E891
AB2300en