Cleaning and care
8.3
Cleaning
Cleaning the device
Cleaning accessories
8.4
Disinfection
26 / 32
■
The water temperature must not exceed 25 °C.
■
To prevent any corrosion due to use of detergents or disinfectants, it is
essential to follow the special application instructions provided by the
manufacturers of the detergent or disinfectant.
Disinfectant:
■
Surface disinfectant (not disinfectant for hands or instruments)
■
Ethanol as the sole active substance.
Do not use an ethanol-propanol mixture to disinfect the viewing window
in the lid of the device.
■
Concentration not less than 30%
■
pH: 6 – 8
■
Non-corrosive
1.
Open the lid.
2.
Switch off the device and disconnect it from the power supply.
3.
Remove accessories.
4.
Clean the centrifuge housing and the centrifuging chamber with soap
or a mild detergent and a damp cloth.
5.
Remove any detergent residues with a damp cloth after using deter-
gents.
6.
The surfaces must be dried immediately after cleaning.
7.
Dry the centrifuging chamber with an absorbent cloth if condensation
forms.
1.
Clean the accessories using the detergent and a damp cloth.
2.
Remove any detergent residues with a damp cloth after using deter-
gents.
3.
Dry the accessories immediately after cleaning using a lint-free cloth
and oil-free compressed air. Dry all cavities completely using oil-free
compressed air.
Disinfection must always be preceded by cleaning the com-
ponents concerned.
See ⮫ Chapter 8.3 'Cleaning' on page 26
Disinfectant concentration and application time according
to the manufacturer's instructions.
Rev.: 12 / 10.2023
AB2300en