5.2
Removing the transport lock
5.3
Setting up and connecting the centrifuge
Setting up the centrifuge
AB2300en
1.
Open the box at the top.
2.
Remove the padding.
3.
Remove the device and accessories by lifting them up out of the box.
4.
Place the device on a stable and level surface.
Personnel:
■
Trained user
The lid is closed.
The mains cable is disconnected from the device.
1
3
1
Fig. 14: Transport lock
1
Spacer sleeves
2
Rubber band
3
Transport lock
4
Screws
1.
Tilt the device on the back of the device.
2.
Unscrew 3 screws (
3.
Remove 3 transport locks (
4.
Keep the screws, spacer sleeves and transport lock in a safe place.
5.
Remove the rubber band (
WARNING
Risk of injury due to failing to maintain a sufficient distance to
the centrifuge.
−
As per EN / IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300 mm around the centrifuge during a centrifugation
run.
−
A distance of 300 mm from the ventilation slots and ven-
tilation openings of the centrifuge must be maintained.
Rev.: 12 / 10.2023
3
1
4
3
4 ) with spacer sleeves ( 1 ).
3 ).
2 ) from the rotor.