Commissioning
5
Commissioning
5.1
Unpacking the centrifuge
16 / 32
The lid is closed.
The mains cable is disconnected from the device.
1
3
1
Fig. 13: Transport lock
1
Spacer sleeves
2
Rubber band
3
Transport lock
4
Screws
1.
Tilt the device on the back of the device.
2.
Insert 3 transport locks (
3.
Screw in 3 screws (
4.
Secure the rotor bucket using a rubber band (
CAUTION
Danger of crushing due to parts falling out of the transport
packaging.
−
Keep the device balanced during the unpacking
process.
−
Only open the packaging at the points provided for this
purpose.
CAUTION
Risk of injury from lifting heavy loads.
−
Provide an adequate number of helpers.
−
Note the weight. See ⮫
on page 9 .
NOTICE
Damage to the device due to improper lifting.
−
Do not lift the centrifuge by the control panel or the
control panel holder.
Personnel:
■
Trained user
Rev.: 12 / 10.2023
3
1
4
3
3 ).
4 ) with spacer sleeves ( 1 ).
Chapter 3.1 'Technical data'
2
2 ).
AB2300en