Disinfecting the device
Disinfecting the accessories
Autoclaving
8.5
Maintenance
Greasing the rubber seal of
the centrifuging chamber
Trunnion greasing
Checking the accessories
Centrifuging chamber damage
inspection
Greasing the motor shaft
Accessories with a limited
service life
AB2300en
CAUTION
Risk of injury due to ingress of water or other liquids.
−
Protect the device against external liquids.
−
Do not disinfect the device using spray.
1.
Open the lid.
2.
Switch off the device and disconnect it from the power supply.
3.
Remove accessories.
4.
Clean the housing and centrifuging chamber using disinfectant.
5.
Remove any disinfectant residues with a damp cloth after using disin-
fectants.
6.
The surfaces must be dried immediately after cleaning.
1.
Disinfect the accessories using the disinfectant.
2.
Wet all cavities with bubble-free disinfectant.
3.
Remove the disinfectant residues or leave them to dry after using
disinfectants.
The rotor and accessories must not be autoclaved.
Rub the sealing ring lightly with a rubber care product.
1.
Remove accessories.
2.
Clean the trunnions.
3.
Remove any detergent residues with a damp cloth after using deter-
gents.
4.
Grease the trunnions and suspension with Hettich Tubenfett 4051.
5.
Excess grease in the centrifuging chamber must be removed.
1.
The accessories shall be checked for wear and corrosion damage.
2.
Check that the rotor is firmly seated.
Check the centrifuging chamber for damage.
1.
Remove accessories.
2.
Clean the motor shaft.
3.
Remove any detergent residues with a damp cloth after using deter-
gents.
4.
Grease the motor shaft with Hettich Tubenfett 4051.
5.
Excess grease in the centrifuging chamber must be removed.
The use of certain accessories is time-limited. For safety reasons, the
accessories must no longer be used when either the maximum number
of permissible run cycles marked on them or the expiry date marked on
them has been reached.