Software operation
7
Software operation
7.1
Centrifugation parameters
7.1.1
Set the brake level
7.1.2
Relative centrifugal force, RCF
Display of the relative centri-
fugal force (RCF)
7.1.3
Centrifugation of substances or mixtures of substances with a density higher than
1.2 kg/dm
28 / 39
1.
Switch off the mains switch.
2.
Press and hold down the ▲
[IMPULSE] button simultaneously.
3.
Switch on the mains switch and release the buttons.
Press the ▲
indicator shows the machine version and the time indicator shows
the brake level set (and/or
The machine version is set ex works and cannot be changed.
The machine version is set ex works and cannot be changed.
4.
Use the
Level 1 = short ramp-down time.
Level 0 = long ramp-down time.
5.
Press the
The relative centrifugal force RCF is dependent on the speed and the centri-
fuging radius.
The relative centrifugal force RCF is stated as a multiple of the acceleration
due to gravity (g).
The relative centrifugal force RCF is a dimensionless numerical value and is
used to compare the separation and sedimentation performance.
RCF = Relative Centrifugal Force
RPM = speed
r = centrifuging radius in mm = distance from the centre of the axis of
rotation to the bottom of the centrifuge tube.
1.
Press and hold the
Relative centrifugal force (RCF) is displayed.
2.
Release the
The speed is displayed.
3
The density of the substances or mixtures of substances must not exceed
1.2 kg/dm³ during centrifugation at maximum speed. The speed must be
reduced for substances or substance mixtures with a higher density. The
permissible speed can be calculated using the following formula:
Rev.: 17 / 01.2024
[RPM/RCF x 100] button and the
[RPM/RCF x 100] button repeatedly until the speed
'0' or '1' ).
[t] buttons to set the desired brake level.
[STOP] button to save the settings.
[RCF] button during the centrifugation run.
[RCF] button.
AB1206en