Télécharger Imprimer la page

Hettich ROTOFIX 32 A Mode D'emploi page 56

Masquer les pouces Voir aussi pour ROTOFIX 32 A:

Publicité

Les langues disponibles

Les langues disponibles

Device overview
3.5.3
Controls
Fig. 6: [Mains switch]
Fig. 7: [IMPULSE] button
Fig. 8: [RCF] button
Fig. 9: [START] button
Fig. 10: [STOP] button
Fig. 11: [RPM/RCFx100] button
Fig. 12: [t] button
3.6
Original spare parts
3.7
Scope of delivery
14 / 39
Switch the device on and off.
Short-time centrifugation. The centrifugation run takes place as long as
the button is being pressed.
Display the brake level and centrifuging radius.
Relative centrifugal force, parameter RCF.
The relative centrifugal force (RCF) is displayed for as long as the button
is pressed.
Start centrifugation run.
End the centrifugation run.
The rotor ramps down to a stop at the preselected brake level.
Save the brake level and centrifuging radius.
A numerical value from 500 RPM to the maximum rotor speed can be
set.
Adjustable in increments of 100 (RPM = displayed value x 100).
Display the brake level and centrifuging radius.
Enter runtime.
Adjustable from 1 to 99 minutes in 1 minute increments.
Centrifuging radius.
Input in centimetres. Adjustable from 5 to 16 centimetres, in 1 centi-
metre increments.
Brake level 0 or 1.
Level 1 = short ramp-down time
Level 0 = long ramp-down time.
Use only original spare parts from the manufacturer and approved accesso-
ries.
The following accessories are supplied with the centrifuge:
1 Release pin
1 hex key (SW5 x 100)
1 grease for the trunnions
Rev.: 17 / 01.2024
AB1206en

Publicité