Commissioning
5.3
Setting up and connecting the centrifuge
Setting up the centrifuge
18 / 39
1
2
3
Fig. 14: Transport lock
1
Transport lock
2
Spacer sleeves
3
Screws
1.
Tilt the device on the back of the device.
2.
Unscrew 3 screws (
3.
Remove 3 transport locks (
4.
Keep the screws, spacer sleeves and transport lock in a safe place.
WARNING
Risk of injury due to failing to maintain a sufficient distance to
the centrifuge.
As per EN / IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300 mm around the centrifuge during a centrifugation
run.
A distance of 300 mm from the ventilation slots and ven-
tilation openings of the centrifuge must be maintained.
CAUTION
Risk of crushing and damage to the device due to it falling
down because of vibration-induced position alterations.
Place the device on a stable and level surface.
Select the installation surface dependent on the weight
of the device.
Rev.: 17 / 01.2024
2
2
3  ) with spacer sleeves ( 2  ).
1  ) and store them safely.
3
2
3