Operation
Removing the hub
4
Fig. 17: Hub installation and
removal
1
Hub
2
Motor shaft
3
Driver
4
Groove
Installing the rotor without a
clamping nut
Installing the hub
Installing the rotor
6.3
Inserting and removing buckets
Inserting buckets
22 / 39
1.
Open the lid.
1
2.
Unscrew the clamping nut.
After passing the working point for lifting the rotor, the hub (
detaches from the cone of the motor (
2
3.
Remove the hub.
3
Personnel:
■
Trained user
1.
Open the lid.
2.
Clean the motor shaft (
3.
Lightly grease the motor shaft (
cleaning and disinfection' on page 31 .
4.
Place the hub (
3 ) of the motor shaft must be in the groove ( 4 ) of the hub.
The driver (
Check that the hub is firmly seated.
5.
Hand-tighten the clamping nut of the hub using the supplied hex key.
6.
Check that the hub is firmly seated.
1.
Clean the hub (
2.
Lift the rotor by the rotary handle and place it vertically on the hub (
3.
Push the rotor down as far as it will go.
NOTICE
Damage to the device due to imbalances caused by incor-
rect loading of the rotor.
−
Buckets marked with the number of the rotor location may
only be used there.
Buckets marked with a set number may only be used
together.
Rev.: 17 / 01.2024
2 ) and rotor hole.
2 ), see ⮫ Chapter 8.2 'Instructions for
1 ) vertically on the motor shaft ( 2 ).
2 ).
Load all swing-out rotor locations with the same
buckets.
1 )
2 ).
AB1206en