8.2 Handling of electromagnetic interactions
Although electromagnetic interference of the secretion aspirator
Allegra M30 has been reduced to a minimum, electromagnetic disturb-
ances between the secretion aspirator Allegra M30 and other
devices cannot be excluded. Therefore you should comply with the specified
requirements and instructions regarding the permissible electromagnetic
environment at any case and monitor the secretion aspirator Allegra
M30 in order to ensure proper operation and to prevent adverse events for
patients and users. Select another location for the secretion aspirator
Allegra M30, if the permissible electromagnetic environment cannot be
ensured or if you have observed functioning of the secretion aspirator
Allegra M30 or of another device in the vicinity that is not intended.
Since electromagnetic propagation is affected by absorption and reflection
from structures, objects and people and since the field strength from fixed
transmitters cannot be predicted in advance with accuracy, a site survey of
the electromagnetic phenomena at the location at which the secretion
aspirator Allegra M30 shall be operated should be considered to assess
the existing electromagnetic environment at this location. If unusual perform-
ance is detected, additional measures may be required, for example a modified
orientation or another location of the secretion aspirator
Allegra M30. If the essential features of the secretion aspirator
Allegra M30 are impaired by electromagnetic disturbances, a reduced or
non-existent flow rate and vacuum may be expected.
All manuals and user guides at all-guides.com
8.3 List of all cables and transformers replaceable by the user
Name
Power
supply unit incl.
connecting cable
Mains cable
Specification
Type: GTM91099-6015-3.0-T2
Manufacturer: GlobTek
Technical data:
100-240 V~, 50-60 Hz, 1.5 A (in)
12 V DC, 5 A (out)
Type: H03VVH2-F
Manufacturer: HAWA
Technical data: 250 V~, 2.5 A
8 EMC information
Maximum length
1.20 m
1.80 m
EN