8 EMC information
Power supply DC
Phenomenon
Test method
Fast transient electrical
IEC 61000-4-4
disturbances/bursts
Impulse voltages/
IEC 61000-4-5
surges
Conducted RF distur-
IEC 61000-4-6
bances
Electrical transient
ISO 7637-2
conduction along
supply lines
EN
78
All manuals and user guides at all-guides.com
Immunity test levels
±2 kV
100 kHz repetition frequency
±1 kV line-to-line
±2 kV line-to-ground
3 V
0.15 MHz to 80 MHz
6V in ISM bands and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
see ISO 7637-2
Patient connection
Phenomenon
Test method
Electrostatic discharge
IEC 61000-4-2
Conducted RF distur-
IEC 61000-4-6
bances
Signal input and output channels
Phenomenon
Test method
Electrostatic discharge
IEC 61000-4-2
Fast transient electrical
IEC 61000-4-4
disturbances/bursts
Impulse voltages/
IEC 61000-4-5
surges
Conducted RF distur-
IEC 61000-4-6
bances
Immunity test levels
±8 kV contact
±15 kV air
3 V
0.15 MHz to 80 MHz
6 V in ISM bands and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Immunity test levels
±8 kV contact
±15 kV air
±1 kV
100 kHz repetition frequency
±2 kV line-to-ground
3 V
0.15 MHz to 80 MHz
6 V in ISM bands and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz